Gauge 3 Specifications

START UP

Calibration to warm-up is 1 minute, after which instrument is fully functional and readings are completely accurate.

DATA LOGGING

Storage for more than 20,000 data points Data intervals from 3 seconds to 15 minutes Memory full display flag.

DISPLAY

Alphanumeric LCD 0.37 inch line readout Two line readout<16 characters per line

MAXIMUM READINGS

Instrument stores maximum pressure. (Maximum pressure is not a peak hold function)

UPDATE INTERVALS

Temperature = 1 minute Barometric = 6 minutes

AUTO SHUT-OFF

Manual engage/disengage (With Auto-Shut-Off engaged, instrument automatically will turn of 15 minutes after last operator input activity.)

POWER SUPPLY

4 "AA" batteries 250 hour maximum service file

CALIBRATION STANDARDS

± 0.0125% of reading or better secondary standards traceable to NIST.

SERIAL INTERFACE

IRDA interface

OVERPRESSURE CAPABILITY

0-2 psi and below 1.5 x range 3-5000 psi 1.5 x range 7,500 psi 1.5 x range

OVERPRESSURE

Displayed at 150% of full scale. (Instruments displaying overpressure error require return to factory to clear mode and verify gaugee accuracy.)

SENSITIVITY

± 0.002% of span with dampening 1 part in 50,000 (maximum)

REPEATABILITY

Ranges 0/5 PSI and above ± 0.01% of span (typical)
Ranges 0/2 PSI and below ± 0.03% of span

SENSORS

Standard Isolated For use with any media compatible with 316 Stainless Steel
Optional Isolated For use with any media compatible with Inconel.

 

PRESSURE RANGES

0-5 0-10 0-15
0-30 0-50 0-100
0-200 0-300 0-500
0-1,000 0-2,000 0-3,000
0-5,000 0-10,000 0-20,000